Design Guidelines for Medical Silicone Components

10/29/20260

Most silicone tooling problems are decided on the drawing, not on the machine. This is the checklist we work through at DFM review, with the reasoning behind each rule so you can apply it before you send the file.

1. Keep Wall Thickness Uniform

Target 1–3 mm for medical LSR parts, and keep the maximum-to-minimum thickness ratio under about 3:1. Thick sections cure more slowly and shrink more than thin ones, which produces sink, warp and inconsistent durometer. Where a thick section is unavoidable, core it out rather than leaving solid material.

2. Radius Every Internal Corner

Inside radii of at least 0.25 mm. Sharp internal corners concentrate stress (tear initiation) and trap air during fill, which shows up as short shots or burn marks.

3. Add Draft to Cored Features

0.5–1° minimum on deep cores and tall walls. Silicone grips tooling, and draft is what lets the part release without distortion or sticking.

4. Decide the Parting Line Early

The parting line leaves a witness mark and potentially flash. Put it where neither matters — never across a sealing face or a cosmetic surface. Say so on the drawing: which faces are sealing, which are cosmetic.

5. Gate in a Non-Critical Area

Gate vestige is a small mark; place it where it will not interfere with function or appearance. For LSR, cold runner systems keep material waste low and hold the mix ratio stable.

6. Design Out Deep Undercuts

Shallow undercuts can often be stripped off the core — roughly up to 15 % of feature depth. Beyond that you are paying for side actions or manual loading, and the money is usually better spent simplifying the geometry.

7. Legible Text and Logos

Minimum height and line width around 0.5 mm in silicone. Fine detail that looks crisp in CAD disappears in an elastomer.

8. Specify Tolerances by Function

Use ISO 3302-1 classes: M1 (precision) for sealing and mating features, M2 for general engineering features, M3 for cosmetic and non-functional dimensions. Calling M1 on every dimension inflates tooling and inspection cost without improving the device. See silicone molding tolerances for the full breakdown.

9. State the Measurement Conditions

Silicone deforms under a gauge. Specify 23 ± 2 °C, at least 24 hours after post-cure, and either limited probe force (commonly ≤ 0.5 N) or non-contact optical measurement. Two labs measuring the same part under different protocols will disagree by more than the tolerance band.

10. Tell Us the Application

Temperature range, media, mating parts, sterilisation method and expected life. That information selects the compound, the durometer and the process — and it is the single most useful thing you can include in an RFQ.

FAQ

What is the thinnest wall you can mould?

Thin walls are possible in LSR, but the limit depends on flow length and fill geometry. Bring the model to DFM review rather than designing to a number from a datasheet.

Can I design a zero-draft part?

Not reliably. Silicone releases by stretching, and zero draft means distortion or sticking on demould. Even 0.5° helps.

Should I specify the compound or just the durometer?

Specify both if you know them. Durometer alone does not fix tear strength, compression set, extractables or biocompatibility documentation.

LSR Injection Molding · Silicone Prototyping

Talk to the engineering team

Send a drawing or a project brief and get a manufacturability review, a process recommendation and a quotation.

Send your drawings for a DFM review


Leave a Reply

Your email address will not be published. Required fields are marked *



Get In Touch


E-mail: [email protected]

86 18106969228

https://lsrinjection.com


Our Network


Room 210, No. 89, Tianan Road, Jimei District, Xiamen, China

Privacy Policy